We are pleased to sponsor the Electrode Placement and Functional Movement series presented by Dr. Lucinda Baker, Associate Professor at USC Division of Biokinesiology and Physical Therapy. These videos provide comprehensive information on the preparation and use of electrodes for effective neurostimulation treatment. All electrodes used in the video presentation are PALS® neurostimulation electrodes. We trust these videos will assist your efforts in providing the most effective treatment to your patients.

DVDs can be ordered by contacting us. Chapter downloads, as well as the full DVD ISO file, are available for free offline viewing via the download links below.

Currently playing:
Chapter 36 [Download]
Full DVD ISO file available: [Download]
PULSE WIDTH: 300uSEC      HZ: 35      WAVEFORM: Asymmetric      STIMULATION GRADE: N/A
Application Instructions:

Electrode placement for flexion reflex, placement two. An asymmetric biphasic waveform is used. The negative electrode is placed just below the fibular head, marked with the X. The positive electrode is placed over the muscle belly of the anterior tibialis muscle. During stimulation you see direct activation of the dorsiflexors with indirect activation of the hip flexors, as indicated by increased knee flexion.